Training and practical issues in gauge block metrology

Project Description

Experts from PTB will give on site training in practical issues of gauge block metrology (wringing etc.). Moreover an examination of different techniques for the determination of phase and roughness correction will be discussed in more detail. The work will be performed at the laboratories of BEV.


Final Report 2007-02-02

This project was dealing with two important aspects of gauge block metrology. The first part was basically an on-site training of BEV-staff and OMH-staff in wringing techniques by experts of the PTB. This part was completed in 2004. The roughness correction for interferometrical gauge block calibration was the main work in thisproject. Different techniques were tested for applicability for the BEV laboratory. The method at the BEV was the simple stack technique which is time consuming and depends on a handful short blocks with nominally the same roughness.

PTB uses the TIS (totally integrating sphere) technique which is fast and convenient but needs a calibration and does not work with ceramic artifacts. During this projectways to calibrate the TIS output were discussed and tested. PTB provided a calibration (together with individual roughness values) of a set of 8 steel gauge blocks. These artifacts were then used at BEV for further work. First a recalibration (with no roughness correction applied) was compared to the values of PTB. Values from the coupled interferometer method were obtained by using transparent wringing platens and inverted mounting. For the evaluations PTB provided measured spectral complex refractive indices of steel. Finally values from the stack technique were also measured.

The results can be seen in the table:

PTB    TIS                                 5 nm ± 01 nm

BEV    coupled interferometer   9 nm ± 10 nm

BEV    stack                               19 nm ± 10 nm

BEV    comparison to PTB         20 nm ± 10 nm
 

Conclusions:

  •  Values obtained by TIS (PTB) and other techniques at BEV do not fit together very well
  • The elegant coupled interferometer technique has problems with the fringe fraction calibration used at BEV (as opposed to phase stepping) and with inhomogeneities in the platen material
  • For the calibrations of a new TIS gauge blocks with different roughness must be available which is not easy
     

Therefore it was decided at BEV to stay at the established stack method.

Subjects
Length (L)
Coordinating Institute
BEV (Austria)
Participating Partners
PTB (Germany)