European comparison in accelaration calibration

Project Description

Campaign to compare the procedures and results, using the laser interferometry method, for the absolute calibration of the medium frequency accelerometers.

Principle: measurement of the longitudinal sensivity

Equipment: - two piezo electric standards according to:

Bruel & Kjaer 8305 (measurement on the top),

Endevco 2270 M8 (measurement on the bottom).

- the same accceleromters in circulation.

Limits : - discrete frequency from 10 Hz to 10 kHz

- discrete g's level from 0,5 to 800 m/s2

Further Information

-Testing: End of the first 1993 semester (round robin).

Schedule: End of the second 1993 semester (report).

-We hope the collaboration of other European countries in particular the one of PTB(DE).

Subjects
Acoustics, Ultrasound and Vibration (AUV)
Coordinator
Michel Fourcade, LNE (France)
Phone: +33 56 68 46 75
Fax: +33 57 71 54 28
Further Partners
C.E.A. - C.E.S.T.A. L.E.A. B.P.2