Bilateral comparison on primary pressure calibration of LS1P and LS2P microphones.

Project Description

The Spanish National Metrology Institute has improved and developed its facilities with a microphone primary calibration system, which was put into functioning in January 2002. In order to check the capability measurement calibration a bilateral comparison is necessary.The frequency range for LS1P microphones is: 20 Hz to 10 kHz, and for LS2P: 20 Hz to 25 kHz.


Final Report 2004-05-17

The bilateral comparison on primary pressure calibration of LS1P and LS2P microphones was carried out between DPLA and CEM. Two microphones from DPLA have been used as transfer standards. Calibrations were performed in 20 Hz to 10 kHz range for LS1P microphones and in 20 Hz to 25 kHz range for LS2P microphones.

The results for LS2P microphone are given in the table below. The table shows the expanded uncertainties (Ua, Ub) as reported by the participants and the degree of equivalence between the two participants is given by a pair of numbers; the difference between the reported calibration results and the associated expanded uncertainty (2÷sqrt(ua^2+ub^2)) both in dB.

Frequency
(Hz)
Expanded
uncertainties in dB
(k=2)
CEM – DPLA
Degree of equivalence
in dB
  DPLA (Ua) CEM (Ub) Difference Exp. Unc.
20 0.1 0.08 -0.065 0.128
31.5 0.08 0.05 -0.056 0.094
63 0.04 0.04 -0.049 0.057
125 0.03 0.04 -0.045 0.05
250 0.03 0.03 -0.041 0.042
500 0.03 0.03 -0.039 0.042
1000 0.03 0.03 -0.038 0.042
1250 0.03 0.03 -0.038 0.042
1600 0.03 0.03 -0.036 0.042
2000 0.03 0.03 -0.037 0.042
2500 0.03 0.03 -0.034 0.042
3150 0.03 0.03 -0.033 0.042
4000 0.03 0.03 -0.035 0.042
5000 0.03 0.03 -0.031 0.042
6300 0.03 0.03 -0.032 0.042
8000 0.03 0.03 -0.033 0.042
10000 0.03 0.03 -0.02 0.042
12500 0.04 0.04 -0.023 0.057
16000 0.05 0.05 0.004 0.071
20000 0.08 0.07 -0.042 0.106
25000 0.14 0.14 -0.098 0.198

 

The results for LS1P microphone are following:

Frequency
(Hz)
Expanded
uncertainties in dB
(k=2)
CEM – DPLA
Degree of equivalence
in dB
  DPLA (Ua) CEM (Ub) Difference Exp. Unc.
20 0.06 0.08 -0.023 0.1
25 0.05 0.05 -0.022 0.071
31.5 0.05 0.05 -0.02 0.071
63 0.03 0.05 -0.016 0.058
125 0.025 0.03 -0.015 0.039
250 0.025 0.03 -0.011 0.039
500 0.025 0.03 -0.009 0.039
1000 0.025 0.03 -0.008 0.039
1250 0.025 0.03 -0.007 0.039
1600 0.025 0.03 -0.012 0.039
2000 0.025 0.03 -0.013 0.039
2500 0.03 0.03 -0.013 0.042
3150 0.03 0.03 -0.016 0.042
4000 0.05 0.03 -0.031 0.058
5000 0.05 0.05 -0.026 0.071
6300 0.05 0.05 -0.035 0.071
8000 0.05 0.07 0.059 0.086
10000 0.08 0.08 0.044 0.113

 

Degree of equivalence between both partners shows that CEM primary pressure microphones calibration system is in reasonably good agreement with DPLA. The observed differences in the calibration results are all within the expanded uncertainties. The results for LS2 microphones will be substituted by the ongoing key-comparison, EUROMET project 674.