News

"Smallest" volume CMC

A new Calibration and Measurement Capability (CMC) was published in the BIPM key comparison data-base in October 2016. The new CMC from IPQ (Portugal) belongs to the field of volume measurements and refers to the calibration of micropipettes using the photometric method (PCS - Pipette Calibration System from Artel). It is so far, the smallest volume capability of any National Metrology Institute (NMI), 10-7 L.

There are two methods used for calibration of micropipettes: The gravimetric method (ISO 8655-6:2002) and the photometric method (ISO 8655-7:2005). The gravimetric method is commonly used by NMIs and Accredited Laboratories to calibrate volume instruments, including micropipettes. This method consists of weighing a delivered volume of the micropipette into a beaker placed on a balance.

The alternative photometric method uses a high-resolution photometer with colorimetric analysis to determine the volume delivered by a micropipette. The basic principle behind photometric measurement is the conservation of mass and the Lambert-Beer Law. In this method, an unknown volume of red dye is delivered into a vial containing a known volume and concentration of blue dye. After mixing, the change in absorbance of the resulting volume can be calculated as a ratio.

In micropipettes the uncertainty component that has a major contribution to the final uncertainty depends on the volume determined. In the photometric method for small volumes the repeatability is the largest uncertainty component but for large volumes, bigger than 100 µL, the instrument will be the most significant source of uncertainty. This is the opposite of the results obtained by the gravimetric method, where the balance is the largest source of uncertainty, and that can get to 30 % of the measured volume for 0,1 µL. In the photometric method the uncertainty for the described volume will be 2 % and therefore the best method to be used for smaller volumes is the photometric method.

Select your area of interest
EMPIR project raises awareness of new calibration methods for electronic brachytherapy devices
2024-04-04

The project partners publicised the metrological developments for electronic brachytherapy and were able to contribute to several standards more

EMPIR project develops procedures needed for traceability in molecular radiotherapy
2024-04-03

EMPIR project established clinical procedures that allow use of traceable methodologies for molecular radiotherapy across Europe more

EMPIR project extends European metrological cohesion in the fight against blindness
2024-04-02

A small collaborative project aimed at extending knowledge in the field of eye tonometers has increased its impact more

Paper describing work of Large Volume Metrology EMPIR project wins outstanding paper award
2024-03-27

Improving large-scale dimensional measurements for manufacturing more

EMPIR project publishes good practice guide to aid uptake of standards related to waveguides
2024-03-25

From 2012 to 2016 the IEEE published standards for high-frequency waveguides. EMPIR NeWITT project supports their uptake more

Page 3 of 227.